DICOM PS3.16 2020a - Content Mapping Resource |
Page 511 |
Note
The concept for syringe (physical object) is used in this context group as distinct from syringe (unit of presentation), which is used in Section CID 68. The concept for bottle is used in the context of consumable used during an oral administration of contrast.
CID 70 Flush
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Extensible |
Version: |
20181115 |
UID: |
1.2.840.10008.6.1.1260 |
Table CID 70. Flush
Coding Scheme |
Code Value |
Code Meaning |
SNOMED-RT ID |
UMLS Concept Unique ID |
Designator |
|
|
|
|
SCT |
13132007 |
Dextran |
C-A7220 |
C0086140 |
SCT |
262003004 |
Saline |
C-70841 |
C0445115 |
SCT |
347379006 |
Lactated Ringer's |
C-70434 |
C0073385 |
CID 71 Imaging Agent Administration Injector Event Type
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Extensible |
Version: |
20181115 |
UID: |
1.2.840.10008.6.1.1261 |
Table CID 71. Imaging Agent Administration Injector Event Type
Coding Scheme Designator |
Code Value |
Code Meaning |
DCM |
130150 |
Pressure above warning limit |
DCM |
130151 |
Pressure above adjustment limit |
DCM |
130152 |
Flow rate above warning limit |
DCM |
130153 |
Flow rate above adjustment limit |
DCM |
130161 |
Keep vein open started |
DCM |
130162 |
Keep vein open ended |
DCM |
130175 |
Air detected |
DCM |
130155 |
Fixed duration pause ended |
DCM |
130163 |
Syringe attached |
DCM |
130164 |
Syringe detached |
DCM |
110501 |
Equipment failure |
DCM |
110527 |
Resource inadequate |
DCM |
130156 |
Terminated due to pressure above termination limit |
DCM |
130157 |
Terminated due to flow rate above termination limit |
DCM |
130176 |
Terminated due to air detected |
DCM |
130158 |
Terminated due to excessive duration pause |
DCM |
130154 |
Terminated due to request from operator |
DCM |
130159 |
Terminated due to injector communication loss |
DCM |
130160 |
Terminated due to unspecified injector failure |
DCM |
130177 |
Terminated by scanner |
Page 512 |
DICOM PS3.16 2020a - Content Mapping Resource |
Coding Scheme Designator |
Code Value |
Code Meaning |
DCM |
130178 |
Terminated due to critical battery level |
DCM |
130179 |
Terminated due to consumable removal |
CID 72 Imaging Agent Administration Step Type |
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Extensible |
|
Version: |
20181115 |
|
UID: |
1.2.840.10008.6.1.1262 |
|
Table CID 72. Imaging Agent Administration Step Type
Coding Scheme Designator |
Code Value |
Code Meaning |
DCM |
130247 |
Patency Test Injection |
DCM |
130248 |
Transit Time Test Injection |
DCM |
130249 |
Diagnostic Administration |
DCM |
130251 |
Flush Administration |
CID 73 Bolus Shaping Curves
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Extensible |
Version: |
20181115 |
UID: |
1.2.840.10008.6.1.1263 |
Table CID 73. Bolus Shaping Curves
Coding Scheme Designator |
Code Value |
Code Meaning |
DCM |
130252 |
Negative exponential |
DCM |
130253 |
Linear Curve |
CID 74 Imaging Agent Administration Consumable Catheter Type
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Extensible |
Version: |
20181115 |
UID: |
1.2.840.10008.6.1.1264 |
Table CID 74. Imaging Agent Administration Consumable Catheter Type
Coding Scheme |
Code Value |
Code Meaning |
SNOMED-RT ID UMLS Concept Unique |
Designator |
|
|
ID |
SCT |
82449006 |
Peripheral intravenous catheter |
SCT |
52124006 |
Central venous catheter |
SCT |
398013009 |
Implantable venous access port |
SCT |
52124006 |
Peripherally inserted central |
|
|
catheter |
SCT |
705541005 |
Rectal Catheter |
A-26836 |
C0179768 |
A-26810 |
C1145640 |
A-1450B |
C1275732 |
A-26810 |
C1145640 |
R-FEAEC |
C0179784 |
CID 75 Low-high-equal
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Non-Extensible |
|
DICOM PS3.16 2020a - Content Mapping Resource |
Page 513 |
Version: |
20181115 |
|
|
|
UID: |
1.2.840.10008.6.1.1265 |
|
|
|
Table CID 75. Low-High-Equal |
|
Coding Scheme |
Code Value |
Code Meaning |
SNOMED-RT ID |
UMLS Concept Unique ID |
Designator |
|
|
|
|
SCT |
62482003 |
Low |
G-A374 |
C0205251 |
SCT |
75540009 |
High |
G-A373 |
C3163633 |
SCT |
9726003 |
Equal |
G-A214 |
C0205163 |
CID 76 Type of Pre-medication |
|
|
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
|
Type: |
Extensible |
|
|
|
Version: |
20181115 |
|
|
|
UID: |
1.2.840.10008.6.1.1266 |
|
|
|
Table CID 76. Type of Pre-Medication |
|
Coding Scheme |
Code Value |
Code Meaning |
SNOMED-RT ID UMLSConceptUnique |
Designator |
|
|
|
ID |
DCM |
123012 |
Pre-Medication |
|
|
DCM |
130259 |
Contrast Reaction Prophylactic |
|
|
|
Agent |
|
|
SCT |
372614000 |
Sedative |
F-6171D |
C0036557 |
SCT |
52017007 |
Antiemetic |
C-85800 |
C0003297 |
CID 82 Units of Measurement
Context Group ID 82 comprises the case-sensitive codes of UCUM. See Section 7.2.2.
Note
Equivalent to the HL7 Value Set "Units of Measure case sensitive" 2.16.840.1.113883.11.12839.
CID 83 Units for Real World Value Mapping
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Extensible |
Version: |
20080123 |
UID: |
1.2.840.10008.6.1.24 |
|
Table CID 83. Units for Real World Value Mapping |
Coding Scheme Designator |
Code Value |
Code Meaning |
Include CID 84 “PET Units” |
|
|
UCUM |
[hnsf'U] |
Hounsfield unit |
CID 84 PET Units
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Extensible |
Version: |
20170914 |
UID: |
1.2.840.10008.6.1.1030 |
Page 514 DICOM PS3.16 2020a - Content Mapping Resource
Table CID 84. PET Units
Coding Scheme Designator |
Code Value |
Code Meaning |
Include CID 85 “SUV Units” |
|
|
UCUM |
{counts} |
Counts |
UCUM |
{counts}/s |
Counts per second |
UCUM |
{propcounts} |
Proportional to counts |
UCUM |
{propcounts}/s |
Proportional to counts per second |
UCUM |
cm2 |
Centimeter**2 |
UCUM |
cm2/ml |
Centimeter**2/milliliter |
UCUM |
% |
Percent |
UCUM |
Bq/ml |
Becquerels/milliliter |
UCUM |
mg/min/ml |
Milligrams/minute/milliliter |
UCUM |
umol/min/ml |
Micromole/minute/milliliter |
UCUM |
ml/min/g |
Milliliter/minute/gram |
UCUM |
ml/g |
Milliliter/gram |
UCUM |
/cm |
/Centimeter |
UCUM |
umol/ml |
Micromole/milliliter |
CID 85 SUV Units
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Extensible |
Version: |
20161106 |
UID: |
1.2.840.10008.6.1.984 |
Table CID 85. SUV Units
Coding Scheme Designator |
Code Value |
Code Meaning |
UCUM |
g/ml{SUVbw} |
Standardized Uptake Value body weight |
UCUM |
g/ml{SUVlbm} |
Standardized Uptake Value lean body mass (James) |
UCUM |
g/ml{SUVlbm(James128)} |
Standardized Uptake Value lean body mass (James 128 |
|
|
multiplier) |
UCUM |
g/ml{SUVlbm(Janma)} |
Standardized Uptake Value lean body mass (Janma) |
UCUM |
cm2/ml{SUVbsa} |
Standardized Uptake Value body surface area |
UCUM |
g/ml{SUVibw} |
Standardized Uptake Value ideal body weight |
Note |
|
|
The formulas for the determination of SUVbw, SUVbsa, SUVlbm (James) and SUVibw are defined in Sugawara et al. Ree- valuation of the Standardized Uptake Value for FDG: Variations with Body Weight and Methods for Correction.Radiology, 1999 at http://radiology.rsna.org/content/213/2/521.
Unfortunately, Sugawara used a parameter of 120 rather than 128 for males, propagating an error in Morgan DJ, Bray KM. Lean Body Mass as a Predictor of Drug Dosage: Implications for Drug Therapy. Clinical Pharmacokinetics. 1994;26(4):292- 307,whichmisquotedtheoriginalLBMdefinitionthatused128inJamesWPT,WaterlowJC.ResearchonObesity:AReport of the DHSS/MRC Group. London: Her Majesty’s Stationery Office; 1976. Implementations differ in whether they have used 120 or 128 for ({SUVlbm}g/ml{SUVlbm}, UCUM, "Standardized Uptake Value lean body mass (James)"). See Kelly M. SUV: Advancing Comparability and Accuracy. Siemens; 2009. Available from: http://www.mpcphysics.com/documents/ SUV_Whitepaper_Final_11.17.09_59807428_2.pdf.
DICOM PS3.16 2020a - Content Mapping Resource |
Page 515 |
The Janmahasatian LBM formula is defined in Janmahasatian et al. Quantification of Lean Bodyweight. Clin Pharmacokinet. 2005 Oct 1;44(10):1051-65. at http://dx.doi.org/10.2165/00003088-200544100-00004 and its role in SUVlbm(Janma) calcu- lation is discussed in Tahari et al. Optimum Lean Body Formulation for Correction of Standardized Uptake Value in PET Imaging.JournalofNuclearMedicine.2014Sep1;55(9):1481-4. athttp://jnm.snmjournals.org/content/55/9/1481.Thepatient size correction factors are summarized here, where weight is in kg and height is in cm:
SUVbw: males & females: weight
SUVlbm (James): males : 1.10 * weight - 120 * (weight/height) ^2
SUVlbm (James): females: 1.07 * weight - 148 * (weight/height) ^2
SUVlbm(Janma): males: 9.27E3 * weight / (6.68E3 + 216 * weight / (height^2))
SUVlbm(Janma): females: 9.27E3 * weight / (8.78E3 + 244 * weight / (height^2))
SUVbsa: males & females: weight^ 0.425 * height^0.725 * 0.007184
SUVibw: males: 48.0 + 1.06 * (height - 152)
females: 45.5 + 0.91 * (height - 152)
CID 91 Functional Condition Present During Acquisition
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Extensible |
Version: |
20170914 |
UID: |
1.2.840.10008.6.1.891 |
Table CID 91. Functional Condition Present During Acquisition
Coding Scheme |
Code Value |
Code Meaning |
SNOMED-RT ID |
UMLS Concept |
Designator |
|
|
|
Unique ID |
Include CID 3271 “Hemodynamic Physiological Challenges” |
|
|
SCT |
43914001 |
Phonation |
F-F7100 |
C0031577 |
SCT |
87731000 |
Weight bearing |
F-12300 |
C0231573 |
DCM |
109137 |
During voiding |
|
|
DCM |
109134 |
Prior to voiding |
|
|
DCM |
109135 |
Post voiding |
|
|
CID 92 Joint Position During Acquisition
Resources: |
HTML | FHIR JSON | FHIR XML | IHE SVS XML |
Type: |
Extensible |
Version: |
20100625 |
UID: |
1.2.840.10008.6.1.892 |
Table CID 92. Joint Position During Acquisition
Coding Scheme |
Code Value |
Code Meaning |
SNOMED-RT ID |
UMLS Concept Unique ID |
Designator |
|
|
|
|
DCM |
109136 |
Neutral musculoskeletal |
|
|
|
|
position |
|
|
SCT |
9964006 |
Flexion |
F-10110 |
C0231452 |
SCT |
24154002 |
Extension |
F-10100 |
C0522009 |
SCT |
60074003 |
Abduction |
F-10120 |
C0231456 |