DICOM PS3.16 2020a - Content Mapping Resource​

Page 1135​

 

NL​

VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

 

 

 

 

Type​

 

11​

NUMERIC​

EV (130092, DCM, "High Dose​

1​

U​

Units=EV(Gy,UCUM,"Gray")​

 

 

 

Fraction Transition Dose")​

 

 

 

12​

NUMERIC​

EV(130093,DCM,"AtomicNumber")​ 1-n​

U​

Units = EV (1, UCUM, "no​

 

 

 

 

 

 

units")​

13​

>​ NUMERIC​

EV (130094, DCM, "Elemental​

1​

M​

Units = EV (ratio, UCUM,​

 

 

 

Composition Atomic Mass Fraction")​

 

 

"ratio")​

14​

NUMERIC​

EV (130095, DCM, "alpha gEUD​

1​

U​

Units = EV (ratio, UCUM,​

 

 

 

value")​

 

 

"ratio")​

Content Item Descriptions​

Rows 12, 13​

Thevalueof(130094,DCM,"ElementalCompositionAtomicMassFraction")annotatesthefractionalweight​

 

of the elements identified by the (130093, DCM, "Atomic Number") with respect to the total mass of the​

 

segment. The allowed value is in the range of [0, 1].​

TID 15302 Patient Support Position Parameters​

Type:​

Extensible​

Order:​

Non-Significant​

Root:​

No​

Table TID 15302. Patient Support Position Parameters​

 

VT​

Concept Name​

VM​

Req​ Condition​ Value Set Constraint​

 

 

 

 

Type​

 

1​

NUMERIC​

EV (126802, DCM, "IEC61217 Table Top​

1​

U​

Units = EV (deg, UCUM,​

 

 

Continuous Pitch Angle")​

 

 

"deg")​

2​

NUMERIC​

EV (126803, DCM, "IEC61217 Table Top​

1​

U​

Units = EV (deg, UCUM,​

 

 

Continuous Roll Angle")​

 

 

"deg")​

3​

NUMERIC​

EV (126801, DCM, "IEC61217 Patient​

1​

U​

Units = EV (deg, UCUM,​

 

 

Support Continuous Yaw Angle")​

 

 

"deg")​

4​

NUMERIC​

EV (126804, DCM, "IEC61217 Table Top​

1​

U​

Units = EV (mm, UCUM,​

 

 

Eccentric Axis Distance")​

 

 

"mm")​

5​

NUMERIC​

EV (126805, DCM, "IEC61217 Table Top​

1​

U​

Units = EV (deg, UCUM,​

 

 

Continuous Eccentric Angle")​

 

 

"deg")​

6​

NUMERIC​

EV (126806, DCM, "IEC61217 Table Top​

1​

U​

Units = EV (mm, UCUM,​

 

 

Lateral Position")​

 

 

"mm")​

7​

NUMERIC​

EV (126807, DCM, "IEC61217 Table Top​

1​

U​

Units = EV (mm, UCUM,​

 

 

Longitudinal Position")​

 

 

"mm")​

8​

NUMERIC​

EV (126808, DCM, "IEC61217 Table Top​

1​

U​

Units = EV (mm, UCUM,​

 

 

Vertical Position")​

 

 

"mm")​

9​

NUMERIC​

EV (126812, DCM, "Isocentric Patient​

1​

U​

Units = EV (deg, UCUM,​

 

 

Support Continuous Pitch Angle")​

 

 

"deg")​

10​

NUMERIC​

EV (126813, DCM, "Isocentric Patient​

1​

U​

Units = EV (deg, UCUM,​

 

 

Support Continuous Roll Angle")​

 

 

"deg")​

11​

NUMERIC​

EV (126814, DCM, "Isocentric Patient​

1​

U​

Units = EV (deg, UCUM,​

 

 

Support Continuous Yaw Angle")​

 

 

"deg")​

12​

NUMERIC​

EV (126815, DCM, "Isocentric Patient​

1​

U​

Units = EV (mm, UCUM,​

 

 

Support Lateral Position")​

 

 

"mm")​

- Standard -​


Page 1136​

 

 

DICOM PS3.16 2020a - Content Mapping Resource​

 

 

 

VT​

 

Concept Name​

 

VM​ Req​ Condition​ Value Set Constraint​

 

 

 

 

 

 

 

Type​

 

13​

NUMERIC​

 

EV (126816, DCM, "Isocentric Patient​

1​

U​

 

Units = EV (mm, UCUM,​

 

 

 

 

Support Longitudinal Position")​

 

 

 

 

"mm")​

14​

NUMERIC​

 

EV (126817, DCM, "Isocentric Patient​

1​

U​

 

Units = EV (mm, UCUM,​

 

 

 

 

Support Vertical Position")​

 

 

 

 

"mm")​

TID 15303 Radiotherapy Treatment Scheduled Processing Parameters​

Type:​

 

 

 

Extensible​

 

 

 

 

 

Order:​

 

 

Non-Significant​

 

 

 

 

 

Root:​

 

 

 

No​

 

 

 

 

 

 

 

Table TID 15303. Radiotherapy Treatment Scheduled Processing Parameters​

 

 

VT​

 

Concept Name​

VM​

Req Type​Condition​ Value Set Constraint​

1​

TEXT​

 

EV (121384, DCM, "RT Plan Label")​

1​

U​

 

 

2​

NUMERIC​

 

EV (121385, DCM, "Current Fraction​

1​

U​

 

UNIT = EV (1, UCUM, "no​

 

 

 

 

Number")​

 

 

 

 

units")​

3​

NUMERIC​

 

EV (121386, DCM, "Number of​

 

1​

U​

 

UNIT = EV (1, UCUM, "no​

 

 

 

 

Fractions Planned")​

 

 

 

 

units")​

4​

NUMERIC​

 

EV (121387, DCM, "Number of​

 

1​

U​

 

UNIT = EV (1, UCUM, "no​

 

 

 

 

Fractions Completed")​

 

 

 

 

units")​

5​

CODE​

 

EV (121388, DCM, "Checked-In​

 

1​

U​

 

DCID 230 “Yes-No”​

 

 

 

 

Status")​

 

 

 

 

 

TID 15304 Radiotherapy Treatment Progress Parameters​

 

 

Type:​

 

 

 

Extensible​

 

 

 

 

 

Order:​

 

 

Non-Significant​

 

 

 

 

 

Root:​

 

 

 

No​

 

 

 

 

 

 

 

 

Table TID 15304. Radiotherapy Treatment Progress Parameters​

 

VT​

 

Concept Name​

VM​

Req Type​

Condition​

Value Set Constraint​

1​

NUMERIC​

EV (121389, DCM, "Referenced​

1​

U​

 

UNIT = EV (1, UCUM, "no​

 

 

 

Beam Number")​

 

 

 

 

units")​

TID 15400 Real-World Quantity Definition​

 

 

 

 

 

Type:​

 

 

 

Extensible​

 

 

 

 

 

Order:​

 

 

Non-Significant​

 

 

 

 

 

Root:​

 

 

 

No​

 

 

 

 

 

 

 

 

 

Table TID 15400. Real-World Quantity Definition​

 

 

NL​

VT​

Concept Name​

VM​

Req Type​Condition​

Value Set Constraint​

1​

 

CODE​

 

DT (246205007, SCT,​

1​

 

M​

BCID 7180 “Abstract​

 

 

 

 

"Quantity")​

 

 

 

Multi-dimensional Image Model​

 

 

 

 

 

 

 

 

Component Semantics”​

2​

 

CODE​

 

BCID 9000 “Physical Quantity​

1-n​

 

U​

 

 

 

 

 

 

Descriptors”​

 

 

 

 

 

Content Item Descriptions​

- Standard -​


 

 

 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 1137​

Row 1​

 

Thisrowusesaconceptnamethatspecifiesthequantifiedcharacteristic.Itisnotrequiredthat(246205007,​

 

 

SCT, "Quantity") be used if there is a reason to use a similar concept.​

 

Row 2​

 

May be concept modifiers, such as (370129005, SCT, "Measurement Method").​

TID 15401 Real-World Quantity Definition for X-Ray Attenuation Properties​

Type:​

 

 

Extensible​

 

 

 

 

Order:​

 

 

Non-Significant​

 

 

 

 

Root:​

 

 

No​

 

 

 

 

 

Table TID 15401. Real-World Quantity Definition for X-Ray Attenuation Properties​

NL​

VT​

Concept Name​

VM​

Req Type​

Condition​

Value Set Constraint​

1​

INCLUDE​

DTID 15400 “Real-World​

1​

M​

 

 

 

 

 

Quantity Definition”​

 

 

 

 

2​

NUMERIC​

EV (DCM, 130087,​

1​

MC​ IF TID 15400 Row 1​

UNITS = EV ("MeV",​

 

 

 

"Reference Energy")​

 

Quantity value is (130086,​UCUM,​

 

 

 

 

 

DCM, "Relative Linear​

"Megaelectronvolt")​

 

 

 

 

 

Stopping Power")​

 

- Standard -​


Page 1138​

DICOM PS3.16 2020a - Content Mapping Resource​

- Standard -​

DICOM PS3.16 2020a - Content Mapping Resource​

Page 1139​

DDICOMControlledTerminologyDefinitions​ (Normative)​

This Annex specifies the meanings of codes defined in DICOM, either explicitly or by reference to another part of DICOM or an ex-​ ternal reference document or standard.​

The contents of this table are available in OWL format at ftp://medical.nema.org/medical/dicom/resources/ontology/dcm/dcm.owl and​

in Bioportal.​

Table D-1. DICOM Controlled Terminology Definitions (Coding Scheme Designator "DCM" Coding​

Scheme Version "01")​

Code Value​

Code Meaning​

Definition​

Notes​

AR​

Autorefraction​

Anacquisitiondevice,processormethodthatmeasures​

 

 

 

autorefraction.​

 

ARCHIVE​

Archive​

A device, process or method that stores images and​

 

 

 

other objects for a prolonged period of time.​

 

AS​

Angioscopy​

An acquisition device, process or method that records​Retired​

 

 

 

images during angioscopy.​

 

ASMT​

Content Assessment Result​

Adevice,processormethodthatproducesassessments​

 

 

 

of the content of other instances, e.g. for quality or​

 

 

 

suitability.​

 

AU​

Audio​

An acquisition device, process or method that records​

 

 

 

audio.​

 

BDUS​

Ultrasound Bone Densitometry​An acquisition device, process or method that performs​

 

 

 

ultrasound bone densitometry.​

 

BI​

Biomagnetic imaging​

An acquisition device, process or method that performs​

 

 

 

biomagnetic imaging.​

 

BMD​

Bone Mineral Densitometry​

An acquisition device, process or method that performs​

 

 

 

bone mineral densitometry by X-Ray, including​

 

 

 

dual-energy X-Ray absorptiometry (DXA) and​

 

 

 

morphometric X-Ray absorptiometry (MXA).​

 

CAD​

Computer Assisted​

An image processing device, process or method that​

 

 

Detection/Diagnosis​

performs computer assisted detection or diagnosis.​

 

CAPTURE​

Image Capture​

An acquisition device, process or method that performs​

 

 

 

image capture, includes video capture.​

 

CD​

Color flow Doppler​

An acquisition device, process or method that performs​Retired​

 

 

 

color flow Doppler.​

 

 

 

Replaced by (US, DCM,​

 

 

"Ultrasound")​

CF​

Cinefluorography​

An acquisition device, process or method that performs​Retired​

 

 

 

cinefluorography.​

 

 

 

Replaced by (RF, DCM,​

 

 

"Radiofluoroscopy")​

COMP​

Computation Server​

A device, process or method that performs computation​

 

 

 

as a service; includes radiotherapy planning.​

 

CP​

Culposcopy​

An acquisition device, process or method that records​Retired​

 

 

 

images during culposcopy.​

 

CR​

Computed Radiography​

An acquisition device, process or method that performs​

 

 

 

computed radiography.​

 

- Standard -​


Page 1140​

DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

CS​

Cystoscopy​

An acquisition device, process or method that records​Retired​

 

 

 

images during cystoscopy.​

 

CT​

Computed Tomography​

An acquisition device, process or method that performs​

 

 

 

computed tomography.​

 

CTPROTOCOL​CT Protocol​

A device, process or method that produces CT device​

 

 

 

acquisition protocols.​

 

DD​

Duplex Doppler​

An acquisition device, process or method that performs​Retired​

 

 

 

duplex Doppler.​

 

 

 

Replaced by (US, DCM,​

 

 

"Ultrasound")​

DF​

Digital fluoroscopy​

An acquisition device, process or method that performs​Retired​

 

 

 

digital fluoroscopy.​

 

 

 

Replaced by (RF, DCM,​

 

 

"Radiofluoroscopy")​

DG​

Diaphanography​

An acquisition device, process or method that performs​

 

 

 

diaphanography.​

 

DM​

Digital microscopy​

An acquisition device, process or method that performs​Retired​

 

 

 

digital microscopy.​

 

DOC​

Document​

A device, process or method that produces documents.​

 

 

 

i.e., representations of documents as images, whether​

 

 

 

by scanning or other means.​

 

DOCD​

Document Digitizer Equipment​A device, process or method that digitizes hardcopy​

 

 

 

documents and imports them.​

 

DS​

Digital Subtraction Angiography​An acquisition device, process or method that performs​Retired​

 

 

 

digital subtraction angiography.​

 

 

 

Replaced by (XA, DCM,​

 

 

"X-Ray Angiography")​

DSS​

Department System Scheduler​A department-based information system (for instance,​

 

 

 

Radiology or Laboratory) that provides functions related​

 

 

 

to the management of orders received from external​

 

 

 

systems or through the department system’s user​

 

 

 

interface. This definition matches that of the DSS/OF​

 

 

 

Actor in the IHE Scheduled Workflow (SWF) Profile.​

 

DX​

Digital Radiography​

An acquisition device, process or method that performs​

 

 

 

digital radiography.​

 

EC​

Echocardiography​

An acquisition device, process or method that performs​Retired​

 

 

 

echocardiography.​

 

 

 

Replaced by (US, DCM,​

 

 

"Ultrasound")​

ECG​

Electrocardiography​

An acquisition device, process or method that performs​

 

 

 

electrocardiography.​

 

EPS​

Cardiac Electrophysiology​

An acquisition device, process or method that performs​

 

 

 

cardiac electrophysiology.​

 

ES​

Endoscopy​

An acquisition device, process or method that records​

 

 

 

images during endoscopy.​

 

F​

Female​

Female sex.​

 

FA​

Fluorescein angiography​

An acquisition device, process or method that performs​Retired​

 

 

 

fluorescein angiography.​

 

 

 

Replaced by (OP, DCM,​

 

 

"Ophthalmic photography")​

FC​

Female changed to Male​

Female sex changed to Male sex.​

 

- Standard -​