DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

113512​

Radiopharmaceutical Lot​

Identifies the vial, batch or lot number from which the​

 

 

Identifier​

individual dispense radiopharmaceutical quantity (dose)​

 

 

 

is produced. The Radiopharmaceutical Dispense Unit​

 

 

 

Identifier records the identification for each individual​

 

 

 

dose.​

 

113513​

Reagent Vial Identifier​

Identifies the lot or unit serial number for the reagent​

 

 

 

component for the radiopharmaceutical.​

 

113514​

Radionuclide Vial Identifier​

Identifiesthelotorunitserialnumberfortheradionuclide​

 

 

 

component for the radiopharmaceutical.​

 

113516​

Prescription Identifier​

Administered Product's Prescription Number​

 

113517​

Organ Dose Information​

Informationpertainingtotheestimatedabsorbedradiation​

 

 

 

dose to an organ.​

 

113518​

Organ Dose​

The absorbed radiation dose to organ​

 

113520​

MIRD Pamphlet 1​

Reference authority​

 

 

 

MIRDPamphletNo.1(rev),SocietyofNuclearMedicine,​

 

 

 

1976​

 

113521​

ICRP Publication 53​

Reference authority​

 

 

 

ICRP, 1988. Radiation Dose to Patients from​

 

 

 

Radiopharmaceuticals. ICRP Publication 53. Ann. ICRP​

 

 

 

18 (1-4).​

 

113522​

ICRP Publication 80​

Reference authority​

 

 

 

ICRP, 1998. Radiation Dose to Patients from​

 

 

 

Radiopharmaceuticals (Addendum to ICRP Publication​

 

 

 

53). ICRP Publication 80. Ann. ICRP 28 (3).​

 

113523​

ICRP Publication 106​

Reference authority​

 

 

 

ICRP, 2008. Radiation Dose to Patients from​

 

 

 

Radiopharmaceuticals-Addendum3toICRPPublication​

 

 

 

53. ICRP Publication 106. Ann. ICRP 38 (1-2).​

 

113526​

MIRDOSE​

Reference authority​

 

 

 

Stabin MG, Sparks RB, Crowe E (1994)​

 

 

 

MIRDOSE:personalcomputersoftwareforinternaldose​

 

 

 

assessment in nuclear medicine [Computer program]​

 

113527​

OLINDA-EXM​

Reference authority​

 

 

 

Stabin MG, Sparks RB, Crowe E (2005) OLINDA/EXM:​

 

 

 

The Second-Generation Personal Computer Software​

 

 

 

for Internal Dose Assessment in Nuclear Medicine​

 

 

 

[Computer program]​

 

113528​

Package Insert​

Reference authority​

 

 

 

The reported organ dose is based on​

 

 

 

radiopharmaceutical's package insert.​

 

113529​

Institutionally Approved​

Reference authority​

 

 

Estimates​

The reported organ dose is based on Institutionally​

 

 

 

 

 

 

approvedestimatesfromtheRadioactiveDrugResearch​

 

 

 

Committee (RDRC) of the institution itself.​

 

- Standard -​


 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 1253​

Code Value​

Code Meaning​

Definition​

Notes​

113530​

Investigational New Drug​

Reference authority​

 

 

 

The reported organ dose is based on an Investigation​

 

 

new drug.​

113540​

Activity Measurement Device​

The type of device that performed the activity​

 

 

measurement.​

113541​

Dose Calibrator​

The device that measures the radiation activity of the​

 

 

radiopharmaceutical​

113542​

Infusion System​

Radiopharmaceutical Infusion System​

113543​

Generator​

Radioisotope Generator​

113550​

Fasting Duration​

The number hours the patient has gone without food.​

113551​

Hydration Volume​

The amount of fluids the patient has consumed before​

 

 

the procedure.​

113552​

Recent Physical Activity​

A description of physical activity the patient performed​

 

 

beforethestartoftheprocedure,suchasthatwhichmay​

 

 

affect imaging agent biodistribution.​

113560​ Acute unilateral renal blockage​Blockage in one of the tubes (ureters) that drain urine​

 

 

from the kidneys​

113561​

Low Thyroid Uptake​

5% or less Thyroid Uptake of Iodine​

113562​

High Thyroid Uptake​

25% or higher Thyroid Uptake of Iodine​

113563​

Severely Jaundiced​

Thepatientexhibitssymptomssevereofjaundiceand/or​

 

 

has a Bilirubin >10 mg/dL.​

113568​

Extravasation visible in image​ Extravasationorparavenousadministrationoftheproduct​

 

 

is visible in the images.​

113570​

Cockroft-Gault Formula​

The measurement method of the Glomerular Filtration​

 

estimation of GFR​

Rate is Cockroft-Gault Formula​

113571​ CKD-EPI Formula estimation of​The measurement method of the Glomerular Filtration​

 

GFR​

Rate is CKD-EPI Formula​

113572​

Glomerular Filtration Rate​

The measurement method of the Glomerular Filtration​

 

(MDRD)​

Rate is MDRD​

113573​

Glomerular Filtration Rate​

The measurement method of the Glomerular Filtration​

 

non-black (MDRD)​

Rate is non-black MDRD​

113574​ Glomerular Filtration Rate black​The measurement method of the Glomerular Filtration​

 

(MDRD)​

Rate is black (MDRD)​

113575​

Glomerular Filtration Rate​

The measurement method of the Glomerular Filtration​

 

female (MDRD)​

Rate is female (MDRD)​

113576​

Glomerular Filtration Rate​

The measurement method of the Glomerular Filtration​

 

Cystatin-based formula​

Rate is Cystatin-based formula​

113577​

Glomerular Filtration Rate​

The measurement method of the Glomerular Filtration​

 

Creatinine-based formula​

Rate is Creatinine-based formula (Schwartz)​

 

(Schwartz)​

 

113601​ Small: < 32.0 cm lateral​

Small body thickness for calcium scoring adjustment.​

 

thickness​

Lateral thickness is measured from skin-to-skin, at the​

 

 

level of the proximal ascending aorta, from an A/P​

 

 

localizer image.​

113602​

Medium: 32.0-38.0 cm lateral​ Medium body thickness for calcium scoring adjustment.​

 

thickness​

Lateral thickness is measured from skin-to-skin, at the​

 

 

level of the proximal ascending aorta, from an A/P​

 

 

localizer image.​

- Standard -​


Page 1254​

DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

113603​ Large: > 38.0 cm lateral​

Large body thickness for calcium scoring adjustment.​

 

 

thickness​

Lateral thickness is measured from skin-to-skin, at the​

 

 

 

level of the proximal ascending aorta, from an A/P​

 

 

 

localizer image.​

 

113605​

Irradiation Event Label​

A human-readable label identifying an irradiation event.​

 

113606​

Label Type​

The type of a human-readable label.​

 

113607​

Series Number​

A number that identifies a Series. Corresponds to​

 

 

 

(0020,0011) in PS3.3.​

 

113608​

Acquisition Number​

A number that identifies an Acquisition. Corresponds to​

 

 

 

(0020,0012) in PS3.3.​

 

113609​

Instance Number​

A number that identifies an Instance. Corresponds to​

 

 

 

(0020,0013) in PS3.3.​

 

113611​

Stationary Acquisition​

Acquisition where the X-Ray source does not move in​

 

 

 

relation to the patient.​

 

113612​

Stepping Acquisition​

Acquisition where the X-Ray source moves laterally in​

 

 

 

relation to the patient.​

 

113613​

Rotational Acquisition​

Acquisition where the X-Ray source moves angularly in​

 

 

 

relation to the patient.​

 

113620​

Plane A​

Primary plane of a Biplane acquisition equipment.​

 

113621​

Plane B​

Secondary plane of a Biplane acquisition equipment.​

 

113622​

Single Plane​

Single plane acquisition equipment.​

 

113630​

Continuous​

Continuous X-Ray radiation is applied during an​

 

 

 

irradiation event.​

 

113631​

Pulsed​

Pulsed X-Ray radiation is applied during an irradiation​

 

 

 

event.​

 

113650​

Strip filter​

Filter with uniform thickness.​

 

113651​

Wedge filter​

Filter with variation in thickness from one edge to the​

 

 

 

opposite edge.​

 

113652​

Butterfly filter​

Filter with two triangular sections.​

 

113653​

Flat filter​

Filter with uniform thickness that is for spectral filtering​

 

 

 

only. E.g., filter out low energy portion of the X-Ray that​

 

 

 

would only contribute to skin dose, but not to image.​

 

113661​

Outline of lobulations​

A polyline defining the outline of a lobulated finding.​

 

113662​ Inner limits of fuzzy margin​

A polyline defining the inner limits of a finding with fuzzy​

 

 

 

margin.​

 

113663​ Outer limits of fuzzy margin​

A polyline defining the outer limits of a finding with fuzzy​

 

 

 

margin.​

 

113664​

Outline of spiculations​

A polyline defining the outline of the spiculations of a​

 

 

 

finding.​

 

113665​

Linear spiculation​

A polyline segment graphically indicating the location​

 

 

 

and direction of a spiculation of a finding.​

 

113666​

Pixelated spiculations​

A collection of points indicating the pixel locations of the​

 

 

 

spiculations of a finding.​

 

113669​

Orthogonal location arc​

Connectedlinesegmentsindicatingthecenteroflocation​

 

 

 

of a finding on an orthogonal view.​

 

113670​ Orthogonal location arc inner​

Connected line segments indicating the inner margin of​

 

 

margin​

the location of a finding on an orthogonal view.​

 

- Standard -​


 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 1255​

Code Value​

Code Meaning​

Definition​

Notes​

113671​

Orthogonal location arc outer​

Connectedlinesegmentsindicatingtheouterlocationof​

 

margin​

a finding on an orthogonal view.​

 

113680​

Quality Control Intent​

This procedure is intended to gather data that is used​

 

 

 

for calibration or other quality control purposes.​

 

113681​

Phantom​

An artificial subject of an imaging study.​

 

113682​

ACR Accreditation Phantom -​ A phantom acceptable for the ACR Computed​

 

 

CT​

Tomography Accreditation program.​

 

113683​

ACR Accreditation Phantom -​ AphantomacceptablefortheACRMagneticResonance​

 

MR​

Imaging Accreditation program.​

 

113684​

ACR Accreditation Phantom -​ A phantom acceptable for the ACR Mammography​

 

 

Mammography​

Accreditation program.​

 

113685​

ACR Accreditation Phantom -​ A phantom acceptable for the ACR Stereotactic Breast​

 

Stereotactic Breast Biopsy​

Biopsy Accreditation program.​

 

113686​

ACR Accreditation Phantom -​ AphantomacceptablefortheACRSPECTAccreditation​

 

ECT​

program (but not for PET).​

 

113687​

ACR Accreditation Phantom -​ A phantom acceptable for the ACR PET Accreditation​

 

 

PET​

program (but not for SPECT).​

 

113688​

ACR Accreditation Phantom -​ A SPECT phantom with a PET faceplate acceptable for​

 

ECT/PET​

both the ACR SPECT and PET Accreditation programs.​

113689​

ACR Accreditation Phantom -​ A PET faceplate (made to fit an existing flangeless or​

 

 

PET Faceplate​

flanged ECT phantom) acceptable for the ACR PET​

 

 

 

Accreditation program.​

 

113690​

IEC Head Dosimetry Phantom​A phantom used for CTDI measurement in head modes​

 

 

according to IEC 60601-2-44, Ed.2.1 (Head 16 cm​

 

 

 

diameter Polymethyl methacrylate PMMA).​

 

113691​

IEC Body Dosimetry Phantom​A phantom used for CTDI measurement in body modes​

 

 

according to IEC 60601-2-44, Ed.2.1 (Body 32cm​

 

 

 

diameter Polymethyl methacrylate PMMA).​

 

113692​

NEMA XR21-2000 Phantom​

A phantom in accordance with NEMA standard​

 

 

 

XR-21-2000.​

 

113701​

X-Ray Radiation Dose Report​ X-Ray Radiation Dose Report.​

 

113702​

Accumulated X-Ray Dose Data​X-Ray dose data accumulated over multiple irradiation​

 

 

events. E.g., for a study or a performed procedure step.​

113704​

Projection X-Ray​

Imaging using a point X-Ray source with a diverging​

 

 

 

beam projected onto a 2 dimensional detector.​

 

113705​

Scope of Accumulation​

Entity over which dose accumulation values are​

 

 

 

integrated.​

 

113706​

Irradiation Event X-Ray Data​

X-Ray dose data for a single Irradiation Event.​

 

113710​

Niobium or Niobium compound​Material containing Niobium or a Niobium compound​

Retired. Replaced by​

 

 

 

(767776000,SCT,"Niobium")​

113711​

Europium or Europium​

Material containing Europium or a Europium compound​Retired. Replaced by​

 

compound​

 

(767775001, SCT,​

 

 

 

"Europium")​

113720​

Calibration Protocol​

Describes the method used to derive the calibration​

 

 

 

factor.​

 

113721​

Irradiation Event Type​

Denotes the type of irradiation event recorded.​

 

113722​

Dose Area Product Total​

Total calculated Dose Area Product (in the scope of the​

 

 

including report).​

 

- Standard -​


Page 1256​

DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

113723​

Calibration DateTime​

Last calibration DateTime for the integrated dose meter​

 

 

or dose calculation.​

 

113724​ Calibration Responsible Party​ Individual or organization responsible for calibration.​

 

113725​

Dose (RP) Total​

TotalDoserelatedtoReferencePoint(RP).(inthescope​

 

 

of the including report).​

 

113726​ Fluoro Dose Area Product Total​Total calculated Dose Area Product applied in​

 

 

 

FluoroscopyModes(inthescopeoftheincludingreport).​

113727​ Acquisition Dose Area Product​TotalcalculatedDoseAreaProductappliedinAcquisition​

 

Total​

Modes (in the scope of the including report).​

 

113728​ Fluoro Dose (RP) Total​

DoseappliedinFluoroscopyModes,relatedtoReference​

 

 

Point (RP). (in the scope of the including report).​

 

113729​ Acquisition Dose (RP) Total​

DoseappliedinAcquisitionModes,relatedtoReference​

 

 

Point (RP). (in the scope of the including report).​

 

113730​

Total Fluoro Time​

TotalaccumulatedclocktimeofFluoroscopyinthescope​

 

 

of the including report (i.e., the sum of the Irradiation​

 

 

 

Duration values for accumulated fluoroscopy irradiation​

 

 

events).​

 

113731​ Total Number of Radiographic​Accumulated Count of exposure pulses (single or​

 

 

Frames​

multi-frame encoded) created from irradiation events​

 

 

 

performed with high dose (acquisition).​

 

113732​

Fluoro Mode​

Mode of application of X-Rays during Fluoroscopy.​

 

113733​

KVP​

AppliedX-RayTubevoltageatpeakofX-Raygeneration,​

 

 

in kilovolts; Mean value if measured over multiple peaks​

 

 

(pulses).​

 

113734​

X-Ray Tube Current​

Mean value of applied Tube Current.​

 

113735​

Exposure Time​

CumulativetimethepatienthasreceivedX-Rayexposure​Retired. Replaced by​

 

 

during the irradiation event​

(113824, DCM, "Exposure​

 

 

 

Time").​

113736​

Exposure​

Mean value of X-Ray Current Time product.​

 

113737​ Distance Source to Reference​Distance to the Reference Point (RP) defined according​

 

Point​

to IEC 60601-2-43 or equipment defined.​

 

113738​

Dose (RP)​

Dose applied at the Reference Point (RP).​

 

113739​

Positioner Primary End Angle​ Positioner Primary Angle at the end of an irradiation​

 

 

 

event. For further definition see (112011, DCM,​

 

 

 

"Positioner Primary Angle").​

 

113740​

PositionerSecondaryEndAngle​Positioner Secondary Angle at the end of an irradiation​

 

 

event. For further definition see (112012, DCM,​

 

 

 

"Positioner Secondary Angle").​

 

113742​

Irradiation Duration​

Clock time from the start of loading time of the first pulse​

 

 

until the loading time trailing edge of the final pulse in​

 

 

 

the same irradiation event.​

 

 

 

Note​

 

 

 

LoadingtimeisdefinedinIEC60601-1-3:2008,​

 

 

 

3.37, and described in IEC 60601-2-54:2009,​

 

 

 

203.4.101.3.​

 

113743​

Patient Orientation​

Orientation of the Patient with respect to Gravity.​

 

113744​

Patient Orientation Modifier​

Enhances or modifies the Patient orientation specified​

 

 

 

in Patient Orientation.​

 

- Standard -​