DICOM PS3.16 2020a - Content Mapping Resource​

Page 1261​

Code Value​

Code Meaning​

Definition​

Notes​

113846​

Target Exposure Index​

The target value used to calculate the Deviation Index​

 

 

 

as defined in IEC 62494-1; see PS3.3 definition of​

 

 

 

Exposure Index Macro.​

 

113847​

Deviation Index​

A scaled representation of the accuracy of the Exposure​

 

 

IndexcomparedtotheTargetExposureIndexasdefined​

 

 

in IEC 62494-1; see PS3.3 definition of Exposure Index​

 

 

Macro.​

 

113850​

Irradiation Authorizing​

The clinician responsible for determining that the​

 

 

 

irradiatingprocedurewasappropriatefortheindications.​

113851​

Irradiation Administering​

Thepersonresponsiblefortheadministrationofradiation.​

113852​

Irradiation Event​

An irradiation event is the loading of X-Ray equipment​

 

 

 

caused by a single continuous actuation of the​

 

 

 

equipment's irradiation switch, from the start of the​

 

 

 

loadingtimeofthefirstpulseuntiltheloadingtimetrailing​

 

 

edge of the final pulse. Any automatic on-off switching​

 

 

 

of the irradiation source during the event is not treated​

 

 

as separate events, rather the event includes the time​

 

 

 

between start and stop of irradiation as triggered by the​

 

 

user. E.g., a pulsed fluoro X-Ray acquisition shall be​

 

 

 

treated as a single irradiation event.​

 

113853​

Irradiation Event UID​

Unique Identifier of an Irradiation Event.​

Retired. Replaced by​

 

 

 

(113769, DCM, "Irradiation​

 

 

 

Event UID")​

113854​ Source of Dose Information​

Method by which dose-related details of an Irradiation​

 

 

 

Event were obtained.​

 

113855​

Total Acquisition Time​

Total accumulated acquisition clock time in the scope of​

 

 

the including report (i.e., the sum of the Irradiation​

 

 

 

Duration values for accumulated acquisition irradiation​

 

 

events).​

 

113856​

Automated Data Collection​

Direct recording of data by a relevant system.​

 

113857​

Manual Entry​

Recordingofdatabyahumanoperator,includingmanual​

 

 

transcription of electronic data.​

 

113858​

MPPS Content​

The data is taken from an MPPS SOP Instance.​

 

113859​

Irradiating Device​

A device exposing a patient to ionizing radiation.​

 

113860​ 15cm from Isocenter toward​

15cm from the isocenter towards the X-Ray source; See​

 

Source​

IEC 60601-2-43.​

 

113861​ 30cm in Front of Image Input​

30cm in front (towards the tube) of the input surface of​

 

Surface​

the image receptor; See FDA Federal Performance​

 

 

 

StandardforDiagnosticX-RaySystems§1020.32(d)(3).​

113862​

1cm above Tabletop​

1cm above the patient tabletop or cradle; See FDA​

 

 

 

Federal Performance Standard for Diagnostic X-Ray​

 

 

 

Systems §1020.32(d) (3).​

 

113863​

30cm above Tabletop​

30cm above the patient tabletop of cradle; See FDA​

 

 

 

Federal Performance Standard for Diagnostic X-Ray​

 

 

 

Systems §1020.32(d) (3).​

 

113864​ 15cm from Table Centerline​

15cm from the centerline of the X-Ray table and in the​

 

 

direction of the X-Ray source; See FDA Federal​

 

 

 

Performance Standard for Diagnostic X-Ray Systems​

 

 

 

§1020.32(d) (3).​

 

- Standard -​


Page 1262​

DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

113865​ Entrance exposure to a 4.2 cm​Standard breast means a 4.2 centimeter (cm) thick​ breast thickness​ compressed breast consisting of 50 percent glandular​

and 50 percent adipose tissue.​

 

See Department of Health and Human Services, Food​

 

and Drug Administration. Mammography quality​

 

standards; final rule. Federal Register. Oct. 28, 1997;​

 

68(208):55852-55994; see 900.2(uu).​

113866​ Copied From Image Attributes​Thedataiscopiedfrominformationpresentintheimage​

 

attributes. E.g., dose attributes such as CTDIvol​

 

(0018,9345).​

113867​ Computed From Image​

The data is computed from information present in the​

Attributes​

imageattributes.E.g.,byusingdosimetryinformationfor​

 

the specific irradiating device make and model, applied​

 

to technique information such as KVP and mAs.​

113868​ DerivedFromHuman-Readable​The data is derived from human-readable reports. E.g.,​

 

Reports​

by natural language parsing of text reports, or optical​

 

 

character recognition from reports saved as images by​

 

 

the irradiating device.​

113870​

Person Name​

The name of a specific person.​

113871​

Person ID​

An identification number or code for a specific person.​

113872​

Person ID Issuer​

The organization that issued a Person ID.​

113873​

Organization Name​

The name of an organization.​

113874​ Person Role in Organization​

The role played by a person in an organization.​

113875​ Person Role in Procedure​

The role played by a person in a procedure.​

113876​ Device Role in Procedure​

The role played by a device in a procedure.​

113877​

Device Name​

The name used to refer to a device; usually locally​

 

 

unique.​

113878​

Device Manufacturer​

Manufacturer of a device.​

113879​

Device Model Name​

Model Name of a device.​

113880​

Device Serial Number​

Serial Number of a device.​

113890​

All Planes​

All planes of a multi-plane acquisition equipment.​

113893​

Length of Reconstructable​

The length from which images may be reconstructed​

 

Volume​

(i.e., excluding any overranging performed in a spiral​

 

 

acquisition that is required for data interpolation).​

 

 

Value is distinct from (1113825, DCM, "Scanning​

 

 

Length"), which is the actual length of the table travel​

 

 

during the entire tube loading, according to IEC​

 

 

60601-2-44,andincludesoverranging.Alsodistinctfrom​

 

 

any actual Reconstructed Volume, which may depend​

 

 

on the slice thickness chosen for a particular​

 

 

reconstruction.​

113895​ Top Z Location of​

The Z location that is the top (highest Z value) of the​

 

Reconstructable Volume​

Reconstructable Volume. Specified as the Z component​

 

 

within the Patient Coordinate System defined by a​

 

 

specified Frame of Reference.​

113896​ Bottom Z Location of​

The Z location that is the bottom (lowest Z value) of the​

 

Reconstructable Volume​

Reconstructable Volume. Specified as the Z component​

 

 

within the Patient Coordinate System defined by a​

 

 

specified Frame of Reference.​

- Standard -​


 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 1263​

Code Value​

Code Meaning​

Definition​

Notes​

113897​ Top Z Location of Scanning​

The Z location that is the top (highest Z value) of the​

 

 

Length​

scanning length. Specified as the Z component within​

 

 

 

the Patient Coordinate System defined by a specified​

 

 

 

Frame of Reference.​

 

113898​ Bottom Z Location of Scanning​The Z location that is the bottom (lowest Z value) of the​

 

 

Length​

scanning length. Specified as the Z component within​

 

 

 

the Patient Coordinate System defined by a specified​

 

 

 

Frame of Reference.​

 

113899​

Exposed Range​

The range along the z axis of the total volume irradiated,​

 

 

 

per IEC 60601-2-44, Ed. 3, 203.115(b).​

 

 

 

The start and stop of loading corresponding to the outer​

 

 

 

edgeofthefullwidthhalfmaximumofthefree-in-airdose​

 

 

 

profile for the beam collimation used.​

 

113900​ Dose Check Alert Details​

Report section about cumulative dose alerts during an​

 

 

 

examination.​

 

113901​ DLP Alert Value Configured​

FlagdenotingwhetheraDLPAlertValuewasconfigured.​

 

113902​ CTDIvol Alert Value Configured​Flag denoting whether a CTDIvol Alert Value was​

 

 

 

configured.​

 

113903​

DLP Alert Value​

Cumulative Dose Length Product value configured to​

 

 

 

trigger an alert; see NEMA XR 25-2010 Dose Check​

 

 

 

Standard.​

 

113904​

CTDIvol Alert Value​

Cumulative CTDIvol value configured to trigger an alert;​

 

 

 

see NEMA XR 25-2010 Dose Check Standard.​

 

113905​

Accumulated DLP Forward​

A forward estimate of the accumulated DLP plus the​

 

 

Estimate​

estimatedDLPforthenextProtocolElementGroup;see​

 

 

 

NEMA XR 25-2010 Dose Check Standard.​

 

113906​ Accumulated CTDIvol Forward​Aforwardestimateatagivenlocationoftheaccumulated​

 

 

Estimate​

CTDIvolplustheestimatedCTDIvolforthenextProtocol​

 

 

 

Element Group; see NEMA XR 25-2010 Dose Check​

 

 

 

Standard.​

 

113907​

Reason for Proceeding​

Reason provided for proceeding with a procedure that​

 

 

 

is projected to exceed a configured dose value.​

 

113908​ Dose Check Notification Details​Reportsectionaboutdosenotificationsduringaprotocol​

 

 

 

element.​

 

113909​

DLP Notification Value​

Flag denoting whether a DLP Notification Value was​

 

 

Configured​

configured.​

 

113910​

CTDIvol Notification Value​

Flag denoting whether a CTDIvol Notification Value was​

 

 

Configured​

configured.​

 

113911​

DLP Notification Value​

Dose Length Product value configured to trigger a​

 

 

 

notification for a given protocol element.​

 

113912​

CTDIvol Notification Value​

CTDIvol value configured to trigger a notification for a​

 

 

 

given protocol element.​

 

113913​

DLP Forward Estimate​

A forward estimate of the DLP for the next Protocol​

 

 

 

Element Group; see NEMA XR 25-2010 Dose Check​

 

 

 

Standard.​

 

113914​

CTDIvol Forward Estimate​

A forward estimate of the CTDIvol for the next Protocol​

 

 

 

Element Group; see NEMA XR 25-2010 Dose Check​

 

 

 

Standard.​

 

113921​

Radiation Exposure​

Theamountofionizingradiationtowhichthepatientwas​

 

 

 

exposed.​

 

- Standard -​


Page 1264​

DICOM PS3.16 2020a - Content Mapping Resource​

 

Code Value​

Code Meaning​

Definition​

Notes​

113922​

Radioactive Substance​

Type, amount and route of radioactive substance​

Retired. Replaced by​

 

Administered​

administered.​

(440252007, SCT,​

 

 

 

"Administration of​

 

 

 

radiopharmaceutical").​

113923​

Radiation Exposure and​

Exposuretoionizingradiationandassociatedpreventive​Retired. Replaced by​

 

Protection Information​

measures used to reduce the exposure of parts of the​(73569-6, LN, "Radiation​

 

 

body like lead apron or eye, thyroid gland or gonad​

Exposure and Protection​

 

 

protection.​

Information")​

113930​ Size Specific Dose Estimation​The Size-Specific Dose Estimate is a patient dose​

 

 

 

estimate that takes into account the size of the patient,​

 

 

such as described in [AAPM Report 204] or [AAPM​

 

 

 

Report220]byusinglineardimensionsmeasuredonthe​

 

 

patient or patient images or estimated from patient age.​

113931​

Measured Lateral Dimension​

The side-to-side (left to right) dimension of the body part​

 

 

being scanned (per [AAPM Report 204]).​

 

113932​

Measured AP Dimension​

The thickness of the body part being scanned, in the​

 

 

 

antero-posterior dimension (per [AAPM Report 204]).​

 

113933​

Derived Effective Diameter​

The diameter of the patient at a given location along the​

 

 

Z-axis of the patient, assuming that the patient has a​

 

 

 

circular cross-section (per [AAPM Report 204]).​

 

113934​ AAPM 204 Lateral Dimension​ The Size Specific Dose Estimation is computed using​

 

 

 

Table 1B (32cm phantom) or Table 2B (16cm phantom)​

 

 

of [AAPM Report 204].​

 

113935​ AAPM 204 AP Dimension​

The Size Specific Dose Estimation is computed using​

 

 

 

Table 1C (32cm phantom) or Table 2C (16cm phantom)​

 

 

of [AAPM Report 204].​

 

113936​ AAPM 204 Sum of Lateral and​The Size Specific Dose Estimation is computed using​

 

 

AP Dimension​

Table 1A (32cm phantom) or Table 2A (16cm phantom)​

 

 

of [AAPM Report 204].​

 

113937​ AAPM 204 Effective Diameter​The Size Specific Dose Estimation is computed using​

 

 

Estimated From Patient Age​

Table 1D (32cm phantom) or Table 2D (16cm phantom)​

 

 

using an effective diameter estimated from the patient's​

 

 

age using Table 3 of [AAPM Report 204].​

 

113940​

System Calculated​

Values calculated from other existing parameters.​

 

113941​

In Detector Plane​

A segmented region of the detector surface within the​

 

 

 

irradiated area (but might not be near the center of the​

 

 

detector).​

 

113942​

X-Ray Reading Device​

AdevicethatcreatesdigitalimagesfromX-Raydetectors​

 

 

(Direct, Indirect or Storage).​

 

113943​

X-Ray Source Data Available​ ParametersrelatedtotheX-Raysource(generator,tube,​

 

 

etc) are available to the recording application.​

 

113944​

X-Ray Mechanical Data​

Parameters related to the X-Ray Mechanical System​

 

 

Available​

(Stand,Table)areavailabletotherecordingapplication.​

113945​ X-Ray Detector Data Available​Parameters related to the X-Ray Detector are available​

 

 

to the recording application.​

 

113946​

Projection Eponymous Name​ Describes the radiographic method of patient, tube and​

 

 

detectorpositioningtoachieveawelldescribedprojection​

 

 

or view.​

 

113947​

Detector Type​

Type of Detector used to acquire data. E.g., Images.​

 

113948​

Direct Detector​

Detector that directly transforms the input signal to pixel​

 

 

values.​

 

- Standard -​


 

DICOM PS3.16 2020a - Content Mapping Resource​

Page 1265​

Code Value​

Code Meaning​

Definition​

Notes​

113949​

Indirect Detector​

Detectorthattransformsanintermediatesignalintopixel​

 

 

 

values. E.g., a scintillator-based detector.​

 

113950​

Storage Detector​

Storage detector that stores a signal that is later​

 

 

 

transformed by a reader into pixel values. E.g., a​

 

 

 

phosphor-based detector.​

 

113951​

Film​

Film that is scanned to create pixel values.​

 

113952​

Table Mount​

The cassette/detector is mounted in the patient table.​

 

113953​

Unmounted Detector​

The cassette/detector is not mounted.. E.g., a cassette​

 

 

 

placed underneath the patient.​

 

113954​

Upright Stand Mount​

The cassette/detector is mounted in an upright stand.​

 

113955​

C-Arm Mount​

The cassette/detector is mounted on a c-arm.​

 

113956​

CR/DR Mechanical​

Method of mounting or positioning a CR/DR cassette or​

 

 

Configuration​

detector.​

 

113957​ Fluoroscopy-Guided Projection​An integrated projection radiography system capable of​

 

 

Radiography System​

fluoroscopy.​

 

113958​

Integrated Projection​

A projection radiography system where the X-Ray​

 

 

Radiography System​

detector, X-Ray Source and gantry components are​

 

 

 

integrated and the managing system is able to access​

 

 

 

details of each component.​

 

113959​

Cassette-based Projection​

A projection radiography system where the X-Ray​

 

 

Radiography System​

detector, X-Ray Source and gantry components are not​

 

 

 

integrated. E.g., cassette-based CR and DR systems.​

 

113961​

Reconstruction Algorithm​

Description of the algorithm used when reconstructing​

 

 

 

the image from the data acquired during the acquisition​

 

 

 

process.​

 

113962​

Filtered Back Projection​

An algorithm for reconstructing an image from multiple​

 

 

 

projectionsbyback-projectingthemeasuredvaluesalong​

 

 

 

the line of the projection and filtering the result to reduce​

 

 

 

blurring.​

 

113963​

Iterative Reconstruction​

An algorithm for reconstructing an image from multiple​

 

 

 

projections by starting with an assumed reconstructed​

 

 

 

image,computingprojectionsfromtheimage,comparing​

 

 

 

the original projection data and updating the​

 

 

 

reconstructedimagebaseduponthedifferencebetween​

 

 

 

the calculated and the actual projections.​

 

113964​ At Surface of Patient​

A point at the surface of the patient within the irradiated​

 

 

 

area where the X-Ray beam enters the patient (i.e.​

 

 

 

towards the tube).​

 

113970​ Procedure Step To This Point​ The period of time from the start of a Procedure Step​

 

 

 

until the time point established by the context of the​

 

 

 

reference.​

 

113980​

Water Equivalent Diameter​

The diameter of a cylinder of water having the same​

 

 

 

X-Ray attenuation as the patient for a specified​

 

 

 

reconstructed slice (e.g., as described in [AAPM Report​

 

 

 

220]).​

 

- Standard -​